LEARN IR · STEP 1 · LESSON 3

How does matter interact with radiation?

Absorption, transmission, reflection and scattering: what happens to light and how that becomes a spectrum.

About 16 minutes to read · Prerequisites: lesson 1 e lesson 2 · Updated October 24, 2026

Look at a windowpane. Sunlight passes through almost all of it; a small part is reflected (at night you see your own reflection); and a tiny fraction is absorbed by the glass. Now look at a glass of milk: almost nothing crosses in a straight line, because the fat droplets deflect the light in every direction. An infrared spectrum does this same bookkeeping, only for each wavenumber. Knowing where light can go is what lets you understand what a spectrum actually measures and when to distrust it.

In this lesson you will:
  • distinguish absorption, transmission, reflection and scattering and write the energy balance that links them;
  • relate transmittance and absorbance through the Beer–Lambert law and convert between them;
  • estimate reflection losses at interfaces and understand why ATR works;
  • recognize what scattering and the width of a band say about the sample and its preparation.

1. Four destinations for light

A beam of power I₀ reaches a sample. Each photon can cross it without interacting (transmission), be absorbed (the energy leaves the beam and stays in the matter), be reflected at an interface or be scattered, that is, deflected in another direction (Figure 1). Because energy is conserved, the balance closes:

I₀ = Ir + Ia + It + Is or r + a + t + s = 1

where r, a, t and s are the fractions of the incident power that follow each path.

sampleI₀ (incident beam)reflection (r)absorption (a)the energy stays in the mattertransmission (t)scattering (s)
Figure 1. The four fates of a beam that hits a sample. The thickness of the arrows is only illustrative: the fractions depend on the material, the wavenumber and the measurement geometry. Didactic sketch.

In the infrared, absorption is the one that carries the chemical information (lesson 1): it only happens where the photon energy, hcν̃, matches the difference between two vibrational levels and the selection rule allows the transition. Reflection and scattering depend on the refractive index and on the shape of the sample. They are not “noise”: they add to the signal and, if ignored, are read as if they were absorption.

Key idea. The spectrometer measures how much light stopped reaching the detector at each ν̃. Not all the light that stopped arriving was absorbed.
Explore further: extinction, absorption and where the energy goes

In optics, the total loss by absorption plus scattering is called extinction. In practice, the Beer–Lambert law describes the measured extinction; therefore the “absorbance” of a spectrum is, strictly speaking, an apparent absorbance when there is reflection and scattering (Bohren and Huffman, 1983; Mayerhöfer et al., 2016). The absorbed energy excites a vibrational mode and is, in general, dissipated as heat very quickly (on the order of picoseconds in liquids, by vibrational relaxation); that is why absorption in the mid-IR does not leave the sample “glowing”.

2. Transmittance and absorbance

The spectrometer measures twice: without the sample (the background spectrum, I₀(ν̃), which includes the source, interferometer, detector and the air in the path) and with the sample (I(ν̃)). The ratio, wavenumber by wavenumber, is the transmittance:

T(ν̃) = I(ν̃) / I₀(ν̃)   (%T = 100 · T)

A uniform layer attenuates light exponentially, because each thin slice removes the same fraction of the light that reaches it: I = I₀ e−αl, with α the absorption coefficient (cm⁻¹) and l the path length (cm). In chemistry the decimal logarithm is used, which defines the absorbance and leads to the Beer–Lambert law:

A = −log₁₀ T = log₁₀(I₀/I) = ε · c · l

ε is the molar absorptivity (L mol⁻¹ cm⁻¹), c the concentration (mol L⁻¹) and l the optical path length (cm). Absorbance is proportional to concentration and path length; transmittance is not (Swinehart, 1962).

10Intensity (a.u.)I₀ (background, no sample)I (with sample)(a) The spectrometer measures twice1000Transmittance (%)T = 25 %(b) Transmittance: T(ν̃) = I(ν̃) / I₀(ν̃)0.70AbsorbanceA = 0.60(c) Absorbance: A(ν̃) = −log₁₀ T(ν̃)4000300020001000Wavenumber (cm⁻¹)
Figure 2. How a transmittance spectrum and an absorbance spectrum arise from two measurements. (a) Background (1300 K source, Planck curve in wavenumber) and the background attenuated by three bands; (b) T = I/I₀; (c) A = −log₁₀ T. At the maximum of the band at 1715 cm⁻¹, T = 25 % and A = 0.60. Signals calculated for teaching; they are not measured spectra.
T (%)A = −log₁₀ T
1000
900.046
500.301
36.80.434
101.00
12.0
0.13.0
0204060801000123T = 50 % → A = 0.301T = 10 % → A = 1T = 1 % → A = 2T = 0.1 % → A = 3Transmittance T (%)Absorbance A
Figure 3. The relation A = −log₁₀ T. Near T = 100 % the curve is nearly linear; near T = 0 it shoots up: a drop in T from 1 % to 0.1 % adds one absorbance unit.

Two practical consequences. First: doubling the path length or the concentration doubles A, but squares T (from 25 % to 6.25 %). That is why band heights are compared in absorbance, not in %T, when thicknesses differ. Second: with A = 2, only 1 % of the light reaches the detector; very intense bands look “flattened” at the bottom of the transmittance scale, and noise starts to dominate. In that case, shorten the path, dilute the sample or use ATR. The exact limit depends on the instrument, the detector and the sample.

Explore further: where the Beer–Lambert law stops holding

The law assumes monochromatic radiation, no stray light, a solute without interactions, and negligible or corrected losses by reflection and scattering. In IR these assumptions often fail: (i) finite resolution makes narrow, intense bands look smaller and broader; (ii) stray light limits the maximum measurable absorbance; (iii) in alcohols and acids, hydrogen-bond association changes the shape and absorptivity of the O–H band with concentration; (iv) in thin films, interference fringes appear; (v) near very intense bands, the refractive index varies (anomalous dispersion) and the intensities stop following the simple law (Mayerhöfer et al., 2016). Do not use A as a direct measure of concentration without calibration with standards under the same conditions.

3. Reflection: what happens at interfaces

Whenever light passes from a medium of refractive index n₁ to another of index n₂, part of it is reflected. At normal incidence and without absorption, the reflected fraction (Fresnel formula) is:

r = ( (n₂ − n₁) / (n₂ + n₁) )²

From air (n₁ ≈ 1) to a material of index n, r = ((n − 1)/(n + 1))². The higher the index, the greater the loss. A window has two faces; adding the multiple reflections and assuming no absorption, the maximum transmission is t = (1 − r)/(1 + r).

Materialn (mid-IR, approx.)r per interfacemax. t of a windowTypical use
KBr1.504.0 %92 %windows and transmission pellets
ZnSe2.4317.4 %70 %ATR crystal; windows
Diamond2.4017.0 %71 %ATR crystal
Ge4.0136.1 %47 %ATR crystal (high-index samples)

Theoretical values for normal incidence, without absorption and without antireflection coating. Indices from manufacturers' catalogs (see reference 9); the index varies with wavenumber.

This has two consequences. A high-index crystal loses part of the beam even without absorbing anything. And, in a transmission spectrum, reflections at the faces of the pellet, film or cell lower the whole baseline: it is a general loss, not a band.

Total reflection and ATR

When light goes from a more refractive medium (the crystal, n₁) to a less refractive one (the sample, n₂ < n₁) with an angle of incidence θ greater than the critical angle θc = arcsin(n₂/n₁), the reflection is total. Even so, the electric field does not drop to zero at once at the interface: an evanescent wave penetrates the sample, and its intensity decays exponentially with distance (Figure 4). If the sample absorbs at a certain ν̃, the reflected beam leaves attenuated at that ν̃. This is attenuated total reflection (ATR).

sample (n₂ < n₁)ATR crystal (n₁)normalθevanescent wave(a few µm; grows with λ)beam
Figure 4. Geometry of an ATR crystal in cross-section, with two reflections at the face in contact with the sample. The evanescent wave penetrates a few micrometers; the depth grows with wavelength (lesson 2). Didactic sketch, not to scale.

What makes ATR so practical is that the effective path is small and reproducible: liquids, powders and polymers can be measured without dilution or pellets, as in the ethanol spectrum of the ethanol sheet. The price is that the relative intensities differ from those of a transmission spectrum (bands at low ν̃ become more intense) and depend on the contact of the sample with the crystal.

Explore further: polarization, complex index and distortions

The reflected fraction depends on the polarization (s and p components) and on the angle; the expression above holds for normal incidence. The complex refractive index ñ = n + iκ explains why, near an intense band, specular reflection produces derivative-shaped features (anomalous dispersion; Kramers–Kronig relations) and why ATR spectra need correction when compared with transmission libraries. The penetration depth and the condition for total reflection are treated in Harrick (1967) and Mirabella (1992).

4. Scattering

Scattering deflects light through heterogeneities: particles, bubbles, roughness, domains with a different refractive index. When the particles are much smaller than the wavelength (Rayleigh regime), the scattered intensity grows with the fourth power of the wavenumber:

Isca ∝ ν̃⁴ (particles ≪ λ)

When the size is comparable to λ, the regime is Mie's, more complex (Bohren and Huffman, 1983). In the mid-infrared, λ ranges from 2.5 to 25 µm; so a powder with particles of a few micrometers scatters considerably. In the spectrum, scattering appears as a sloping baseline (the loss increases toward high wavenumbers) and as deformed bands, asymmetric. In KBr pellets, the common practice is to grind the sample until the particles are smaller than the shortest wavelength of interest, precisely to reduce this effect.

Quick test. A = −log₁₀ T does not separate absorption from scattering. A baseline that rises smoothly from right to left (more loss at high ν̃) is a classic sign of scattering; true bands have positions and widths typical of vibrations.
Explore further: the Christiansen effect

Near an intense band, the refractive index of the powder changes rapidly. Where it equals that of the surrounding medium, scattering almost disappears; as a result, in dispersed powders the band can show a characteristic asymmetry on the higher-wavenumber side (Griffiths and de Haseth, 2007). It is one more reason to standardize preparation when comparing spectra.

5. Shape and width of a band

An absorption band is not an infinitely thin line. In condensed samples, the width has two main origins. The first is the finite lifetime of the vibrational states: by the uncertainty principle, a short lifetime implies a poorly defined energy, and the profile is Lorentzian. The second is the distribution of environments of the molecules (hydrogen bonds, conformations): each environment absorbs at a slightly different ν̃, and the sum has a profile close to the Gaussian. Real profiles usually fall between the two (Figure 5), and the instrument resolution is added on top (lesson 12).

full width at half maximum (FWHM)Lorentzianlong tailsGaussianfalls off fastDistance from band center (in units of FWHM)
Figure 5. Lorentzian and Gaussian profiles with the same full width at half maximum (FWHM) and the same maximum. The Lorentzian has long tails, which overlap neighboring bands. Calculated curves.

The O–H band of alcohols is the classic example: the molecules form hydrogen bonds of different strengths, each environment absorbs at a distinct ν̃, and the sum is the broad band near 3300 cm⁻¹ of ethanol. The C=O band of a ketone, in a more uniform environment, is much narrower. The width, therefore, also informs about the environment of the molecule.

Key idea. Position tells which vibration; intensity tells how much (with the caveats of section 2); width and shape tell in what environment. All three enter the reading (lesson 14).
Explore further: width and lifetime

For a state of mean lifetime τ, the full width at half maximum of the Lorentzian profile in energy is ħ/τ. In wavenumber: Δν̃ = 1/(2π c τ), with c = 2.998 × 10¹⁰ cm s⁻¹. With pure dephasing, the width increases, and the expression gives a lower limit. In liquids, the relevant vibrational relaxation times are in picoseconds, which explains widths from a few to tens of cm⁻¹ (Hollas, 2004).

6. Three measurement geometries

GeometryWhat the beam doesTypical preparationWatch for
TransmissionCrosses the sampleKBr pellet, film, liquid cellControlled thickness; scattering in pellets; fringes in films
Reflection (specular or diffuse)Reflects off the surface or the powderSmooth surface or dispersed powderBand shapes distorted by dispersion; dependence on roughness
ATRReflects inside the crystal; the evanescent wave touches the sampleDirect contact with the crystalIntensities depend on λ and on contact; ATR correction to compare with libraries

Lesson 10 compares transmission and ATR in detail. What matters for now is that each geometry favors a different fate of the light, and this changes the spectrum you see.

In practice. When you submit a spectrum to espectroFTIR, state correctly whether the axis is transmittance or absorbance: this information changes how the bands are read.

7. Common misconceptions

8. Worked problems

Each problem applies a lesson section. Try it before opening the solution.

Problem 1 · applies section 2

From transmittance to absorbance

Problem. In a thin film, a band has T = 25 % at its maximum (baseline at 100 %). (a) What is the absorbance? (b) What happens to A and T if the film thickness doubles? (c) And if, instead, the concentration drops to half? (d) What fraction of the light was removed at the maximum of the original band?

View solution
  1. (a) A = −log₁₀(0.25) = 0.602.
  2. (b) With twice the path length, A = 2 × 0.602 = 1.204, and T = 10−1.204 = 0.0625, that is, 6.25 %.
  3. (c) With half the concentration, A = 0.301 and T = 10−0.301 = 50 %.
  4. (d) 1 − 0.25 = 75 % of the incident light no longer reached the detector.

Answer. A = 0.60; with twice the thickness, A = 1.20 and T = 6.25 %; with half the concentration, A = 0.30 and T = 50 %; 75 % of the light was removed. If the losses by reflection and scattering are negligible, those 75 % correspond to absorption; otherwise, the absorption is smaller.

Problem 2 · applies section 3

When diamond gives total reflection

Problem. A diamond crystal (n₁ = 2.40) measures an organic sample of n₂ = 1.50 at an angle of incidence of 45°. (a) What is the critical angle? (b) Is there total reflection? (c) What is the largest sample refractive index for which total reflection still occurs at 45°? (d) What fraction of the light is reflected at the air/diamond face at normal incidence, and what is the maximum transmission of a diamond window?

View solution
  1. (a) θc = arcsin(n₂/n₁) = arcsin(1.50/2.40) = arcsin(0.625) = 38.7°.
  2. (b) 45° > 38.7°: total reflection occurs.
  3. (c) Total reflection requires sin 45° > n₂/n₁, so n₂ < 2.40 × 0.7071 = 1.70.
  4. (d) r = ((2.40 − 1)/(2.40 + 1))² = 0.1696, about 17.0 %; t = (1 − r)/(1 + r) = 0.710, about 71 %.

Answer. θc ≈ 38.7°; total reflection occurs; samples with n₂ above ≈ 1.70 do not produce it at 45° with diamond (that is what germanium, with n ≈ 4, is used for); r ≈ 17 % per face and max. t ≈ 71 % for a window.

Problem 3 · applies sections 4 and 5

Width from lifetime and Rayleigh scattering

Problem. (a) A vibrational state has a mean lifetime of 1 ps. What is the minimum full width at half maximum of the band, in cm⁻¹? (b) And if the mean lifetime is 0.1 ps? (c) For particles much smaller than λ, how many times more intense is Rayleigh scattering at 3000 cm⁻¹ than at 1000 cm⁻¹?

View solution
  1. (a) Δν̃ = 1/(2π c τ) = 1/(2π × 2.998 × 10¹⁰ cm s⁻¹ × 10⁻¹² s) = 5.31 cm⁻¹.
  2. (b) With τ = 10⁻¹³ s, Δν̃ = 53.1 cm⁻¹ (ten times larger).
  3. (c) Isca ∝ ν̃⁴, so the ratio is (3000/1000)⁴ = 3⁴ = 81.

Answer. ≈ 5.3 cm⁻¹ and ≈ 53 cm⁻¹; scattering is 81 times greater at 3000 cm⁻¹ (valid only in the Rayleigh regime). The first value is a lower limit: dephasing, different environments and the instrument resolution only increase the observed width.

9. Questions to consider

  1. A band has A = 3. What is the transmittance and why is this measurement unreliable?
    Answer

    T = 10⁻³ = 0.1 %. Almost no light reaches the detector: noise and stray light start to dominate and the value of A no longer reflects the real absorption. Reduce the path length or the concentration.

  2. Why do bands at low wavenumbers look relatively more intense in ATR than in transmission?
    Answer

    The penetration depth of the evanescent wave is proportional to λ (lesson 2): at lower ν̃ the wave penetrates deeper into the sample and the effective path is longer.

  3. In a transmission spectrum of a poorly ground powder, at which end does the baseline drop most?
    Answer

    At the high-wavenumber end, because Rayleigh scattering grows with ν̃⁴. As the particles approach λ, the regime is Mie's, and the profile can be more complicated.

  4. Why are band heights in %T not compared between two samples of different thicknesses?
    Answer

    Because T depends exponentially on the path length (T = 10−A); only A is proportional to thickness. Convert to absorbance before comparing.

10. Lesson summary

Next lesson: energy levels and transitions (lesson 4).

11. Further reading

Good starting points: Atkins, Stuart and Hecht. To go further: Griffiths and de Haseth, Mirabella, and Bohren and Huffman.

  1. Atkins, P.; de Paula, J.; Keeler, J. Atkins’ Physical Chemistry. 11th ed. Oxford: Oxford University Press, 2018. Supports: absorption, the Beer–Lambert law and line broadening (sections 1, 2 and 5).
  2. Griffiths, P. R.; de Haseth, J. A. Fourier Transform Infrared Spectrometry. 2nd ed. Hoboken: Wiley-Interscience, 2007. Supports: background spectrum, transmittance, scattering and Christiansen effects (sections 2 and 4).
  3. Stuart, B. H. Infrared Spectroscopy: Fundamentals and Applications. Chichester: Wiley, 2004. Supports: measurement geometries and overview (section 6).
  4. Hecht, E. Optics. 5th ed. Boston: Pearson, 2016. Supports: Fresnel formulas, total reflection and the evanescent wave (section 3).
  5. Harrick, N. J. Internal Reflection Spectroscopy. New York: Interscience (Wiley), 1967; Mirabella, F. M. (ed.). Internal Reflection Spectroscopy: Theory and Applications. New York: Marcel Dekker, 1992. Support: condition for total reflection and penetration depth (section 3).
  6. Bohren, C. F.; Huffman, D. R. Absorption and Scattering of Light by Small Particles. New York: Wiley, 1983 (reprinted 1998). DOI 10.1002/9783527618156. Supports: extinction, Rayleigh and Mie scattering (sections 1 and 4).
  7. Swinehart, D. F. The Beer-Lambert Law. Journal of Chemical Education, vol. 39, p. 333, 1962. DOI 10.1021/ed039p333. Supports: formulation of the law (section 2).
  8. Mayerhöfer, T. G.; Mutschke, H.; Popp, J. Employing Theories Far beyond Their Limits: The Case of the (Boguer-) Beer–Lambert Law. ChemPhysChem, vol. 17, pp. 1948–1955, 2016. DOI 10.1002/cphc.201600114. Supports: limits of the law and apparent absorbance (sections 1 and 2).
  9. Bruker Optics. Technical material on ATR crystals (diamond, ZnSe and Ge), accessed on October 18, 2026. Supports: catalog values of the refractive indices in the table of section 3 (they vary with wavenumber).
  10. Hollas, J. M. Modern Spectroscopy. 4th ed. Hoboken: Wiley, 2004. Supports: lifetime and linewidth (section 5).

Edition, pages and content of the reference works must be checked directly; the list records the source of each statement and does not replace reading the original text. For the two articles and for the book by Bohren and Huffman, only the metadata were checked on October 24, 2026 (Crossref).

Original teaching material by espectroFTIR. Diagrams and signals calculated for teaching; they do not represent measured spectra.