LEARN IR · STEP 1 · LESSON 2

What is light?

Waves, wavenumber and the energy scale of spectroscopy.

About 14 minutes to read · Prerequisites: basic waves and quantized energy · Updated October 18, 2026

Infrared spectrometers rarely use glass windows. Ordinary glass transmits visible light but absorbs much of the mid-infrared; this is why their windows and crystals are made of KBr, ZnSe or diamond. To understand why, and almost everything that follows in spectroscopy, we need to understand light.

In this lesson you will:
  • understand light as a wave and as a packet of energy (photon);
  • use the λ, ν and ν̃ scales and convert between them;
  • locate the mid-infrared in the electromagnetic spectrum and understand why it reveals vibrations;
  • understand, without heavy mathematics, how FTIR separates the “colors” of light.

1. Light is a wave

Shake one end of a rope and a wave travels along it. In light, an electric field oscillates together with a magnetic field, perpendicular to each other and to the direction of propagation (Figure 1). In a vacuum, light travels at c = 299 792 458 m s⁻¹.

λ E₀ z E: electric field (vertical plane) B: magnetic field (orthogonal plane) k: direction of propagation (z) E and B oscillate in phase, with B₀ = E₀/c, and are perpendicular to each other and to k.
Figure 1. Linearly polarized plane wave. The wavelength λ is the distance between two consecutive maxima of a field, measured along z. Teaching diagram in oblique projection.

Three quantities describe the wave: wavelength λ (the distance between two maxima), frequency ν (the number of oscillations per second) and speed c, which relates them:

c = λ · ν

When light enters a material, its speed and wavelength change; its frequency does not change. See problem 2 at the end of this lesson.

Explore further: plane waves, irradiance and polarization

For a monochromatic plane wave propagating along z and linearly polarized: E(z,t) = E₀ cos(kz − ωt + φ), with B₀ = E₀/c, k = 2π/λ, ω = 2πν and ω/k = c, and c = 1/√(ε₀μ₀). The average irradiance in a medium of refractive index n is I = ½ n c ε₀ E₀². Polarization describes the trajectory of E in the transverse plane (linear, circular or elliptical). The globar in spectrometers emits essentially unpolarized, broadband radiation; polarization becomes important again in reflectance and ATR (s and p components).

In a medium, v = c/n and λ = λ₀/n. In spectroscopy, the axis is defined in vacuum: ν̃ = 1/λ₀ does not change when light passes through an ATR crystal, a window or the sample.

2. Three scales for the same light

Light can be described by wavelength, frequency or wavenumber (ν̃ = 1/λ, the number of waves per centimeter). In infrared spectroscopy we use wavenumber because it is proportional to energy and independent of the medium.

QuantitySymbolCommon unitRelationship
Wavelength (in vacuum)λnm, µmλ = c/ν
FrequencyνHz (THz)ν = c/λ = c·ν̃
Wavenumberν̃cm⁻¹ν̃ = 1/λ = ν/c
Angular frequencyωrad s⁻¹ω = 2πν
Angular wavenumberkrad m⁻¹k = 2π/λ = 2πν̃
Photon energyEJ, eV, kJ mol⁻¹E = hν = hcν̃
ν̃ (cm⁻¹) = 10 000 / λ (µm)

Example: radiation at 5 µm corresponds to 10 000/5 = 2000 cm⁻¹. A shorter wavelength means a higher wavenumber and higher energy.

Key idea. cm⁻¹ is not frequency in Hz: it is wavenumber, proportional to frequency (ν̃ = ν/c). Informal texts often use “frequency in cm⁻¹” as shorthand, but the correct term is preferable.
Explore further: exact constants and useful conversions

Using the exact SI values (h = 6.626 070 15 × 10⁻³⁴ J s; c = 299 792 458 m s⁻¹; kB = 1,380 649 × 10⁻²³ J K⁻¹; NA = 6.022 140 76 × 10²³ mol⁻¹):

1 cm⁻¹ corresponds to 1.986 446 × 10⁻²³ J, 0.123 984 meV, 11.9627 J mol⁻¹ and 29.9792 GHz
1 eV corresponds to 8065.54 cm⁻¹ · kBT/(hc) = 0.695 036 cm⁻¹ K⁻¹ × T (≈ 207 cm⁻¹ at 298 K)

Also, ν (Hz) = 2.998 × 10¹⁰ · ν̃ (cm⁻¹).

3. Light also consists of packets

When absorbed or emitted, light behaves as packets of energy called photons. The energy of each photon is:

E = hν = hcν̃

Imagine rain: the energy of each drop is its “color” (wavenumber), while the number of drops per second is the intensity. More intense light does not have more energetic photons; it has more photons.

Explore further: detectors and the semiclassical treatment

Thermal detectors (such as pyroelectric DTGS) respond to absorbed power; quantum detectors (such as photoconductive MCT) respond to the photon flux above an energy threshold. IR absorption is treated semiclassically: the field is classical and the molecule is quantum mechanical. A transition is possible when ΔE = hcν̃ and the transition moment ⟨f|μ̂·E|i⟩ is nonzero; this second condition gives rise to the selection rule in lesson 7.

4. Mapping the electromagnetic spectrum

Figure 2 shows the regions on a logarithmic wavenumber scale. Notice two things: the mid-infrared (4000 to 400 cm⁻¹) occupies a narrow range, and it lies well above ambient thermal energy.

RadioMicrowavesFIRUVX-raysMIRNIRVisible10⁻³10⁻²10⁻¹10⁰10¹10²10³10⁴10⁵10⁶10⁷10⁸10⁹Wavenumber (cm⁻¹), logarithmic scalekBT (298 K) ≈ 207 cm⁻¹1 eV ≈ 8066 cm⁻¹Wavelength:1 m1 mm10 µm1 µm100 nm10 nm0.1 nm
Figure 2. Spectral regions on a logarithmic wavenumber scale. The mid-infrared (4000 to 400 cm⁻¹) spans only one decade of energy and lies well above thermal energy at room temperature. Boundaries are conventional; see the text.
Regionλ (approx.)ν̃ (cm⁻¹)Energy per photonTypical phenomenon
X-rays0.01–10 nm10⁶–10⁹124 eV–124 keVinner-shell electron transitions
Ultraviolet10–400 nm2,5 × 10⁴–10⁶3.1–124 eVvalence electronic transitions; photoionization
Visible≈ 400–780 nm25 000–12 8003.1–1.6 eVelectronic transitions
Near-infrared (NIR)0,78–2,5 µm12 800–40001.6–0.50 eVovertones and combination bands (mainly X–H)
Mid-infrared (MIR)2,5–25 µm4000–4000.50–0.050 eVfundamental vibrations and combination bands
Far-infrared (FIR)25–1000 µm400–1050–1.2 meVlow-frequency vibrations, lattice modes, rotations of light molecules
Microwaves1 mm–1 m10–0,011.2 meV–1.2 µeVmolecular rotations; electron paramagnetic resonance
Radiofrequency> 1 m< 0.01< 1.2 µeVnuclear magnetic resonance
Why IR reveals vibrations. At room temperature, thermal motion corresponds to about 207 cm⁻¹. Molecular vibrations (400 to 4000 cm⁻¹) have much higher energies; therefore almost all molecules are in the ground state, and the bands we observe are transitions from the ground state to the first excited level.
Explore further: thermal populations, hot bands and conventional boundaries

The gap between the first vibrational levels of typical bonds is about 400 to 4000 cm⁻¹, or 5 to 48 kJ mol⁻¹. For a mode at 1000 cm⁻¹, the population ratio between v = 1 and v = 0 is exp(−1000/207) ≈ 0.8%. Modes below about 400 cm⁻¹ have appreciable excited-state thermal populations, giving rise to hot bands.

A note on boundaries. Regional boundaries are conventional. Optics and photonics standards (such as ISO 20473) use different mid- and far-infrared boundaries from those common in analytical chemistry, where MIR = 4000 to 400 cm⁻¹ and, in many texts, NIR extends to about 12 500 cm⁻¹. When reporting results, state the range actually measured.

5. How FTIR uses overlapping waves

When two waves of the same frequency meet, they add together: they can reinforce or cancel each other depending on the path difference. This is interference, the basis of the FTIR spectrometer.

The instrument splits the beam in two, sends one beam along a variable path (moving mirror), and recombines them. The detector records the interferogram (Figure 3); the computer then converts it into a spectrum using a Fourier transform. Rather than scanning one “color” at a time, FTIR measures them all simultaneously.

path difference δ(a) Monochromatic radiationpath difference δ(b) Interferogram (broadband)(c) Spectrum B(ν̃)wavenumber ν̃ (cm⁻¹)04000FT
Figure 3. (a) Monochromatic radiation at 2000 cm⁻¹: the interferogram is a cosine with period 1/ν̃ = 5 µm. (b) Broadband source (Gaussian centered at 2000 cm⁻¹, standard deviation 300 cm⁻¹): the cosines cancel far from δ = 0, creating the central burst. (c) Spectrum recovered by Fourier transform. Signals calculated for teaching purposes.
Key idea. A pure color produces a cosine in the interferogram; several colors produce a sum of cosines that reinforce each other only near δ = 0 (the central burst). The Fourier transform separates the sum and recovers the spectrum.
Explore further: the interferometer equation, resolution and coherence

In an ideal Michelson interferometer (50:50 beamsplitter), with monochromatic radiation of wavenumber ν̃ and intensity I₀, detector intensity as a function of optical path difference δ is I(δ) = ½ I₀ [1 + cos(2π ν̃ δ)]. For polychromatic radiation, the variable part of the signal is the Fourier cosine transform of the spectrum B(ν̃): I(δ) = ∫ B(ν̃) cos(2π ν̃ δ) dν̃.

  • Without apodization, resolution is of the order Δν̃ ≈ 1/δmax: 4 cm⁻¹ requires δmax ≈ 0.25 cm, or about 1.25 mm of mirror travel, because the path difference is twice the displacement.
  • The coherence length of a broadband thermal source, lc ≈ 1/Δν̃, is a few micrometers; therefore fringes occur only near δ = 0.
  • A He–Ne laser (about 633 nm) provides a scale for measuring δ accurately.

FTIR details are covered in step 3 (lesson 9).

6. When light passes through matter

What happens depends on wavenumber and material: light can be absorbed, transmitted, reflected or scattered. Absorption follows the Beer–Lambert law (A = εcl). In ATR, light undergoes total internal reflection in the crystal and “escapes” only as an evanescent wave that penetrates a few micrometers into the sample, penetrating deeper at longer wavelengths. These ideas return in lessons 3, 10 and 11.

Explore further: complex refractive index and penetration depth

The complex refractive index ñ(ν̃) = n(ν̃) + iκ(ν̃) summarizes optical properties: κ is related to absorption (α = 4πκν̃), while n varies near an intense band (anomalous dispersion). n and κ are linked by the Kramers–Kronig relations; thus strong bands in reflectance can appear distorted or derivative-shaped.

In ATR, the penetration depth is dp = λ / [2π n₁ √(sin²θ − (n₂/n₁)²)], where λ is the vacuum wavelength, n₁ and n₂ are the crystal and sample refractive indices, and θ is the angle of incidence. Since dp ∝ λ, bands at lower wavenumbers are relatively more intense in ATR than in transmission.

7. Common misconceptions

8. Worked problems

Each problem applies a section of the lesson. Useful constants: c = 2.998 × 10¹⁰ cm s⁻¹; 1 eV = 8065.5 cm⁻¹; 1 cm⁻¹ corresponds to 11.963 J mol⁻¹; kBT/(hc) = 207.2 cm⁻¹ at 298 K (section 2). Try solving the problems before opening the solutions.

Problem 1 · applies sections 2 and 4

The carbonyl band on energy scales

Problem. The C=O band of an aliphatic ketone appears at 1715 cm⁻¹. Calculate: (a) vacuum wavelength; (b) frequency and period; (c) photon energy in eV and kJ mol⁻¹; (d) the ratio of this energy to kBT at 298 K and the approximate fraction of molecules in the first excited vibrational level relative to the ground state.

View solution
  1. (a) λ = 10⁴/1715 = 5,83 µm.
  2. (b) ν = c·ν̃ = 2.998 × 10¹⁰ cm s⁻¹ × 1715 cm⁻¹ = 5.14 × 10¹³ Hz; the period is 1/ν = 1.9 × 10⁻¹⁴ s, about 19 fs.
  3. (c) E = 1715/8065.5 = 0.2126 eV; in molar energy, 1715 × 11.963 = 20 517 J mol⁻¹ ≈ 20.5 kJ mol⁻¹.
  4. (d) E/kBT = 1715/207.2 = 8.28; the population ratio is exp(−8.28) ≈ 2.5 × 10⁻⁴.

Answer. λ = 5.83 µm; ν = 5.14 × 10¹³ Hz (≈ 19 fs); E = 0.213 eV ≈ 20.5 kJ mol⁻¹, or 8.3 kBT. About 0.03% of molecules occupy the first excited level: almost all are in the ground state, so the observed band is the fundamental transition.

Problem 2 · applies section 1 (change of medium)

What changes when radiation enters an ATR crystal

Problem. Radiation at 1000 cm⁻¹ (λ₀ = 10 µm in vacuum) passes through a diamond crystal with refractive index n ≈ 2.4 in the mid-infrared. Determine, inside the crystal: (a) wavelength; (b) phase velocity; (c) frequency. Does the spectral wavenumber change?

View solution
  1. (a) λ = λ₀/n = 10/2,4 = 4,2 µm.
  2. (b) v = c/n = 2,998 × 10⁸ m s⁻¹ / 2,4 = 1,25 × 10⁸ m s⁻¹.
  3. (c) Frequency is set by the source and conserved at the interface: ν = c·ν̃ = 2.998 × 10¹⁰ × 1000 = 3.0 × 10¹³ Hz.

Answer. Inside the crystal, λ ≈ 4.2 µm and v ≈ 1.25 × 10⁸ m s⁻¹; frequency (3.0 × 10¹³ Hz) does not change. Spectral wavenumber, defined in vacuum (ν̃ = 1/λ₀ = 1000 cm⁻¹), also remains unchanged: the spectral axis therefore does not depend on the crystal used.

Problem 3 · applies section 5

Spectral resolution and mirror travel

Problem. An FTIR spectrometer measures at 4 cm⁻¹ resolution. (a) What must the maximum optical path difference, δmax, be? (b) How far must the moving mirror travel? (c) What happens to this displacement if resolution improves to 2 cm⁻¹?

View solution
  1. (a) Without apodization, Δν̃ ≈ 1/δmax; therefore δmax ≈ 1/4 cm⁻¹ = 0.25 cm = 2.5 mm.
  2. (b) The path difference is twice the mirror displacement (light travels there and back): displacement = δmax/2 = 1.25 mm.
  3. (c) For 2 cm⁻¹, δmax = 0.5 cm and mirror travel becomes 2.5 mm.

Answer. δmax ≈ 0.25 cm and mirror travel ≈ 1.25 mm for 4 cm⁻¹; halving the resolution value (to 2 cm⁻¹) doubles the required travel (2.5 mm).

9. Questions to consider

1. Two photons have wavenumbers of 3300 cm⁻¹ and 1100 cm⁻¹. What are the ratios of their energies and wavelengths?

The energy ratio is 3:1 because E = hcν̃. The wavelength ratio is 1:3 because λ = 1/ν̃: the 3300 cm⁻¹ photon has λ ≈ 3.03 µm and the 1100 cm⁻¹ photon has λ ≈ 9.09 µm.

2. Why is the IR spectral axis marked in cm⁻¹ rather than µm?

Because ν̃ is linear in energy and frequency, allowing direct comparison of bands with vibrational energy gaps and theoretical calculations, and because its vacuum definition is independent of the refractive index of the medium. The µm scale is nonlinear in energy (E ∝ 1/λ).

3. Why is almost the entire population of a 3000 cm⁻¹ mode in the ground state at 298 K?

Because hcν̃/(kBT) = 3000/207 ≈ 14.5; the population ratio is exp(−14.5) ≈ 5 × 10⁻⁷. For low-frequency modes (for example, 200 cm⁻¹), the ratio reaches about 0.4 and hot bands become relevant.

4. Why does a broadband thermal source produce interference fringes only near δ = 0?

Because the cosines for each ν̃ have different periods and quickly fall out of phase; the coherence length is lc ≈ 1/Δν̃. For a bandwidth of hundreds of cm⁻¹, lc is a few micrometers and the interferogram is concentrated around zero path difference.

10. Lesson summary

Next lesson: how matter interacts with radiation (lesson 3).

11. Further reading

Good starting points: Atkins, Stuart and Hecht. For further study: Griffiths and de Haseth, and Hollas.

  1. Tiesinga, E.; Mohr, P. J.; Newell, D. B.; Taylor, B. N. CODATA recommended values of the fundamental physical constants: 2018. Reviews of Modern Physics, vol. 93, art. 025010, 2021. DOI 10.1103/RevModPhys.93.025010. Supports: values of h, c, kB, NA and conversion factors (sections 2 and 4).
  2. Griffiths, P. R.; de Haseth, J. A. Fourier Transform Infrared Spectrometry. 2nd ed. Hoboken: Wiley-Interscience, 2007. Supports: the Michelson interferometer, interferogram, resolution and coherence (section 5).
  3. Stuart, B. H. Infrared Spectroscopy: Fundamentals and Applications. Chichester: Wiley, 2004. Supports: infrared regions and an overview of the technique (section 4).
  4. Atkins, P.; de Paula, J.; Keeler, J. Atkins’ Physical Chemistry. 11th ed. Oxford: Oxford University Press, 2018. Supports: electromagnetic radiation, quantization, thermal populations and selection rules (sections 1 to 4).
  5. Hollas, J. M. Modern Spectroscopy. 4th ed. Hoboken: Wiley, 2004. Supports: spectral regions and types of transition (section 4).
  6. Hecht, E. Optics. 5th ed. Boston: Pearson, 2016. Supports: plane waves, polarization, interference and coherence (sections 1 and 5).
  7. ISO 20473:2007. Optics and photonics: Spectral bands. Geneva: ISO. Supports: differences between optical standards and the boundaries used in analytical chemistry (section 4).

Consult the original editions, pages and content directly; this list records the source for each statement and does not replace reading the original text.

Original teaching material by espectroFTIR. Diagrams and signals calculated for teaching; they do not represent measured spectra.